Solid-phase extraction of a carcinogen, 4,4'-methylenedianiline, in serum

H Shintani1

  • 1National Institute of Hygienic Sciences, Department of Medical Devices, Tokyo, Japan.

Insights

This study presents an effective solid-phase extraction method for removing the carcinogen 4,4'-methylenedianiline (MDA) from medical device materials. The optimized pretreatment ensures accurate analysis of MDA in blood serum, enhancing medical device safety.

Area of Science:

  • Analytical Chemistry
  • Materials Science
  • Toxicology

Background:

  • Polyurethane (PU) is widely used in medical devices.
  • Sterilization methods like gamma-ray and autoclaving degrade PU, producing the carcinogen 4,4'-methylenedianiline (MDA).
  • Accurate detection of MDA in biological samples is crucial for assessing health risks.

Purpose of the Study:

  • To develop and optimize a solid-phase extraction (SPE) method for the pretreatment of MDA in blood serum.
  • To evaluate the efficacy of various SPE columns for MDA recovery.
  • To identify the predominant binding mechanisms in the SPE process.

Main Methods:

  • Solid-phase extraction (SPE) using Bond Elut C18, phenyl, cyclohexyl, C8, C1, C2, silica, and SCX columns.
  • Analysis of MDA in spiked human blood serum samples.
  • Elution with methanol containing 1M NH4OH.

Main Results:

  • 100% recovery of MDA from serum was achieved using C18, phenyl, and cyclohexyl columns at concentrations of 0.1-10 ppm.
  • Bond Elut C8 showed 90% recovery, while C1, C2, and silica columns yielded unsatisfactory results.
  • SCX columns provided satisfactory recovery but required complex sample pretreatment and column conditioning.

Conclusions:

  • C18, phenyl, and cyclohexyl SPE columns are highly effective for MDA pretreatment in serum, suggesting pi-pi and van der Waals interactions are key.
  • The developed SPE method enhances the accurate quantification of MDA, contributing to improved safety assessments of medical devices.