Relationship between endotracheal tube leakage and under-reading of tidal volume in neonatal ventilators
R A Mahmoud1, H S Fischer, H Proquitté
1Sohag Faculty of Medicine, Sohag University, Egypt.
Aim:
Protective ventilation in neonates requires careful volume monitoring to prevent ventilator-induced lung injury caused by baro/volutrauma and hence chronic lung disease. This study investigated the effect of endotracheal tube (ET) leakage on the displayed tidal volume using an in vitro model.
Methods:
A neonatal lung model was ventilated via a 3 mm ET using three ventilators [Babylog 8000 (BL), Leoni (LE) and Stephanie (ST)]. Tidal volume was measured by each ventilator at the Y-piece and by a pneumotach (CO(2)SMO(+)) in the model. ET leaks were simulated by open tubes of different lengths. PIP (20 cmH(2)O) and PEEP (5 cmH(2)O) were kept constant, and the respiratory rate (RR) was varied between 20/min and 70/min (Ti:Te = 1:1).
Results:
Tidal volume displayed by a ventilator decreased independently of RR with increasing leakage up to 21% (BL), 30% (LE) and 33% (ST). However, the volume delivered to the lung was nearly constant. The displayed leakage varied between 0 and 78% and was dependent on RR and leakage resistance. There were distinct differences between the three ventilators in the relationship between displayed leakage and volume error. Accepting a volume error <10% for RR between 20 and 70/min, ET leakage of up to 20% for BL, 12% for LE, but only <5% for ST, was acceptable.
Conclusion:
Tidal volume underestimation arising from ET leakage depends on ventilator pressures, timing parameters and ventilator-specific algorithms for signal processing. Therefore, neonatologists should be aware of these issues to prevent lung over-inflation when adjusting target volume in the presence of ET leakage.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Mechanical Ventilation I: Indication and Settings
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...