1-Minute quality control tests for positron emission tomography radiopharmaceuticals
Ryuji Nakao1, Takehito Ito, Kazutaka Hayashi
1Molecular Imaging Center, National Institute of Radiological Sciences, 4-9-1, Anagawa, Inage-ku, Chiba 263-8555, Japan. nakaor@nirs.go.jp
Abstract:
An ultra-fast, sensitive and versatile radio-liquid chromatographic (LC) procedure was developed and validated for quality control (QC) tests of PET radiopharmaceuticals. For a wide variety of radio-probes, the usual LC conditions were used: (1) column: Waters XBridge RP(18) (50 mm x 3.0 mm ID, 2.5 microm), (2) mobile phase: a mixture of three modifiers (90% CH3CN, ammonium phosphate at pH 2.1 and pH 9.3), and (3) detection: UV absorption and NaI(Tl) scintillation. The introduction of a short column packed with small particles of 2.5 microm allowed excellent separation of target analytes within a very short run time of 1 min; only a 3% decline of radioactivity was observed during QC analysis of (11)C-labelled pharmaceuticals. Combining ammonium-phosphate buffer as the mobile-phase component and low-wavelength UV detection led to an improvement in the applicability and sensitivity. All 34 pharmaceuticals investigated could be successfully applied to determine the specific radioactivity, radiochemical and chemical purity with 10-times better sensitivity than traditional LC. We could analyze different pharmaceuticals in a short period since this system utilized a common column and mobile phase. The proposed procedure fulfils the requirements for routine QC tests in terms of rapidity, sensitivity, simplicity and applicability.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...