A high-throughput LC-MS/MS method for the quantitation of posaconazole in human plasma: Implementing fused core
Jennifer M Cunliffe1, Carl F Noren, Roger N Hayes
1Schering-Plough Research Institute, Summit, NJ 07901, United States.
Abstract:
A rapid and robust liquid chromatographic tandem mass spectrometric (LC-MS/MS) method for the determination of posaconazole concentrations in human plasma was validated. Posaconazole was extracted from human plasma using mixed-mode cation exchange solid phase extraction in a 96-well plate format followed by gradient separation on a fused-core Halo C18 column. The analyte and its corresponding internal standard were detected using a Sciex API 4000 triple quadrupole LC-MS/MS system equipped with a TurboIonSpray ionization source operated in the positive ion mode. The calibration range of the method was 5.00-5000ng/mL using a 50microL aliquot of plasma. The assay inter-run accuracy and precision were-4.6-2.8% and 2.3-8.7%, respectively (n=18). The results from method validation indicate the method to be sensitive, selective, accurate, and reproducible. The method was successfully applied to the routine analysis of clinical samples with the fused-core silica columns providing excellent reproducibility for greater than 1000 injections per column.