Purpose:
The accuracy of monitors for measuring transcutaneous PCO2 (TcPCO2), end-tidal PCO2 (EtPCO2), and nasal EtPCO2 was evaluated.
Methods:
The measuring devices included a TcPCO2 monitor (TCM3; Radiometer Trading), an EtPCO2 monitor (Ultima; Datex-Ohmeda), and a nasal EtPCO2 monitor (TG-920P; Nihon Kohden). The sensor electrode of the TCM3 TcPCO2 monitor was applied to the skin of the subject's upper arm. A sampling tube attached to the proximal end of the tracheal tube was connected to the Ultima EtPCO2 monitor. The miniature sensor of the TG-920P nasal EtPCO2 monitor was attached to the nostril. The values obtained were compared with direct measurements of arterial PCO2 (PaCO2) obtained by means of an ABL700 blood gas analyzer (Radiometer Trading) in surgically treated patients. The means +/- 2 SD of the differences between variables were calculated.
Results:
The TcPCO2 monitor (0.19 +/- 4.8 mmHg, mean +/- 2-SD) was more accurate than the EtPCO2 monitor (-4.4 +/- 6.5 mmHg, mean +/- 2-SD) in patients receiving artificial ventilation via an endotracheal tube and the TcPCO2 monitor was also more accurate than the nasal EtPCO2 monitor (-6.3 +/- 9.8 mmHg, bias +/- 2-SD) in patients breathing spontaneously.
Conclusion:
We found that the TcPCO2 monitor was more accurate than the EtPCO2 or nasal EtPCO2 monitor in surgically treated patients.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History