Is the physical decay correction of the (18)F-FDG input function in dynamic PET imaging justified?
Eric Laffon1, Olivier Barret, Roger Marthan
1Service de Médecine Nucléaire, CHU de Bordeaux, Pessac, France. elaffon@u-bordeaux2.fr
Unlabelled:
In this theoretic note, the rationale for the physical decay correction of the (18)F-FDG input function in dynamic PET is investigated, using the Patlak equation as an example.
Methods:
The Patlak equation conventionally obtained when correcting the (18)F-FDG input function and correcting the tissue activity measurement for (18)F physical decay can also be derived from a 2-compartment analysis that does not conceptually involve any physical decay correction of the (18)F-FDG input function but accounts only for the physical decay of the trapped tracer.
Results:
We demonstrate that exactly the same equation can be derived from the 2 conceptual approaches, and hence each approach yields the correct uptake rate of the tracer.
Conclusion:
No advantage in (18)F-FDG dynamic PET can be expected from using the concept of uncorrected data rather than that of decay-corrected data. Nevertheless, conceptually, we show that correcting the (18)F-FDG input function for radioactive decay cannot be justified and that this correction is not compatible with the calculation of patient radiation dose.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET