Pediatric self-inflating resuscitators: the dangers of improper setup
James O'Neill1, Charry Scott, Niranjan Kissoon
1Department of Emergency Medicine, University of Florida at Jacksonville, Jacksonville, Florida, USA.
Background:
Self-inflating resuscitators (SIRs) are often used in pediatric resuscitation. Improper setup of the SIR can lead to inadequate ventilation and oxygenation.
Objectives:
To present clinical scenarios in which SIRs delivered inadequate tidal volumes due to improper use. Second, to evaluate tidal volumes delivered using SIRs at varying lung compliances with the manometer and pop-off valve port open and closed. Third, to suggest methods to overcome improper use.
Methods:
Five pediatric resuscitators were tested under conditions simulating normal lungs, lungs with moderately decreased compliance, and lungs with poor compliance (0.015, 0.010, and 0.005 L/cm H(2)O, respectively) to determine volumes delivered with proper SIR setup (manometer and pop-off valve closed) and improper SIR setup (manometer or pop-off valve open).
Results:
With each SIR, an open manometer port or an open pop-off valve (improper setup) led to significant decreases in volume delivered. In normal lung compliance, the proper setup delivered 149 ± 10 cc, vs. 112 ± 12 cc, 106 ± 25 cc, and 90 ± 14 cc (pop-off open, manometer open, and both open, respectively). In poor lung compliance, the proper setup delivered 122 ± 13 cc, vs. 56 ± 10 cc, 70 ± 17 cc, and 44 ± 7 cc (pop-off open, manometer open, and both open, respectively). All differences above are significant (p < 0.0001).
Conclusions:
In a normal lung, the volumes delivered by SIRs are significantly decreased with the pop-off valve and manometer port open. Proper set-up of the SIR becomes even more important when lung compliance is poor.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation I: Adult
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation I: Indication and Settings