The use of high-resolution digital imaging technology for small diameter K-file length determination in endodontics
Bart Vandenberghe1, Marius Bud, Asti Sutanto
1Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Katholieke Universiteit Leuven, Kapucijnenvoer 7, Leuven, Belgium. bart.vandenberghe@med.kuleuven.be
Abstract:
To assess the reliability of high resolution intra-oral photostimulable storage phosphor (PSP) and complementary metal-oxide semiconductor (CMOS) imaging systems for working length (WL) assessment of small K-files in narrow and curved root canals. Eleven narrow and curved canals from extracted molars were used as pre-test for sample-size calculation. Nineteen canals from four cadavers were used for endodontic length assessment in the final study. Small K-files (ISO size 6, 8, and 10) were introduced into the canals at prepared length. Digital intra-oral radiographs were obtained using high-resolution Vistascan PSP plates and Sigma M CMOS active pixel sensor with a DC X-ray tube at 70 kV, 7 mA, and 0.16 s. Both image series were assessed with and without use of a dedicated endodontic filter. Three observers measured WLs for comparison to the gold standards of a digital millimeter ruler. Multiple regression analysis of the dependent measurements revealed no significant influence of imaging sensor (PSP or CMOS, p = 0.34) and image processing (p = 0.97). For ISO file size, however, there was a significant difference (p = 0.08) at a level of 10%. Observers mostly underestimated lengths using PSP but overestimated them on CMOS. Almost all radiographic measurements (96-98%) were within 2-mm deviation, while 71% to 82% deviated within 1 mm. Dedicated filtering and sensor type did not influence the outcome of WL determination of small file sizes when using high-resolution imaging sensors. WL determination with ISO file 6 did show a significant difference compared to ISO 8 and 10 but mostly for deviations <1.5 mm.