Related Experiment Video
Updated: Jun 23, 2026

Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
Published on: July 14, 2023
A randomised crossover trial comparing the i-gel supraglottic airway and classic laryngeal mask airway
C Janakiraman1, D B Chethan, A R Wilkes
1Department of Anaesthetics and Intensive Care medicine, University Hospital of Wales, Heath Park, Cardiff, UK. karthichitra@gmail.com
Abstract:
In a randomised cross-over study, we compared the performance of the single use i-gel supraglottic airway and reusable classic laryngeal mask airway (cLMA) in 50 healthy anaesthetised patients who were breathing spontaneously. Primary outcome was successful insertion at first attempt. Secondary outcomes included overall insertion success rate, ease of insertion, leak pressure and fibreoptic position. Success rate for insertion at the first attempt was significantly different (54% with i-gel vs 86% with cLMA; p = 0.001). Overall success after two attempts (when the anaesthetist was allowed to change the size of the device) improved to 84% with i-gel vs 92% with cLMA; p = 0.22. In 14 patients, the i-gel when used first needed to be replaced with a larger size. Leak pressure was higher for the i-gel (median [IQR] 20 [14-24] cm H(2)O than the cLMA 17 [12-22] cm H(2)O; p = 0.023). The fibreoptic view through the device was significantly better with the i-gel than the cLMA, which was statistically significant (p = 0.03). We conclude that, with its current sizing recommendations, the i-gel is not an acceptable alternative to cLMA. However because of the significantly improved success rate after a larger sized i-gel was used, we recommend the manufacturer to review the sizing guidelines to improve the success rate.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Cardiopulmonary Resuscitation II: ACLS Airway Management