Related Experiment Video
Updated: Jun 22, 2026

Use of Capillary Aerosol Generator in Continuous Production of Controlled Aerosol for Non-Clinical Studies
Published on: April 12, 2022
Quantitative analysis and design of a spray aerosol inhaler. Part 1: effects of dilution air inlets and flow paths
P Worth Longest1, Michael Hindle
1Department of Mechanical Engineering, Virginia Commonwealth University, Richmond, Virginia., Department of Pharmaceutics, Virginia Commonwealth University, Richmond, Virginia. pwlongest@vcu.edu
Background:
The objective of this study was to evaluate the effects of modifying inhaler design variables on aerosol drug deposition within the mouthpiece for a representative spray system using a quantitative analysis and design approach.
Methods:
Capillary aerosol generation (CAG) was selected as a model spray aerosol system in conjunction with four prototype inhaler body and mouthpiece combinations. In vitro experiments were used to determine drug deposition in the mouthpiece and induction port. Validated computational fluid dynamics (CFD) simulations were implemented to establish relationships between design variables, transport characteristics, and aerosol drug deposition.
Results:
Results of this study indicated that both the size of the upstream dilution air inlets and the flow pathway configuration near the spray nozzle significantly influenced aerosol transport and deposition. CFD results showed that the primary transport characteristics associated with drug deposition were turbulence intensity and the effective diameter of the mouthpiece. Strong quantitative correlations were developed between the identified transport characteristics and mouthpiece drug deposition.
Conclusions:
Based on quantitative analysis and design, turbulence intensity and effective mouthpiece diameter were identified as key transport characteristics within the design space that directly influenced aerosol deposition and may be used to predict and optimize drug delivery to the patient.
Related Concept Videos
Pipe Flowrate Measurement: Problem Solving
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
Inhaled Medications
Application of Integration: Problem Solving
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below: