Development and application of the condom catheter method for non-invasive measurement of bladder pressure
R van Mastrigt1, J J M Pel, J W N C Huang Foen Chung
1Sector Furore, Department of Urology, Erasmus MC, The Netherlands.
Objectives:
A non-invasive method to measure the bladder pressure in males using a condom catheter has been developed. The measurement technique, its validation and limitations, a diagnostic nomogram to non-invasively diagnose bladder outlet obstruction (BOO), and results of large-scale application are discussed.
Methods:
Modified incontinence condoms are attached to the penis. During voiding the flow of urine is mechanically interrupted. The subsequent maximum pressure in the condom reflects the isovolumetric bladder pressure. The method was validated in a group of 46 patients with lower urinary tract symptoms who were simultaneously studied invasively and non-invasively. Subsequently it was applied in a non-invasive epidemiological study in 1020 healthy males.
Results:
The reproducibility of the measured isovolumetric bladder pressure is comparable to that of conventional pressure-flow parameters. The measured pressure can be used to diagnose bladder outlet obstruction with a diagnostic accuracy (Area Under receiver operator characteristic curve) of 0.98, which compares most favorably with the area under the curve of 0.79 of Q(max) in the same population. During condom catheter measurements, both the involuntary interruption of voiding and the forced diuresis increase post-void residual volume. This increase does not affect the accuracy of the pressure measurements.
Conclusions:
We conclude that in males bladder pressure can successfully be measured non-invasively using the condom catheter method. By combining the measured volumetric bladder pressure with a separately measured free flow rate, BOO can non-invasively and accurately be diagnosed.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Urodynamic Studies: Uroflowmetry
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)