Comparison of hemodialysis access flow measurements using flow dilution and in-line dialysance
William L Whittier1, Hansen A Mansy, Daniel R Rutz
1Department of Internal Medicine, Division of Nephrology, Rush University Medical Center, Chicago, Illinois 60607, USA. william_whittier@rush.edu
Abstract:
Measurement of hemodialysis (HD) access flow (QA) is a noninvasive approach for arteriovenous graft or fistula surveillance. Flow dilution (FD) and in-line dialysance (DD) are two common methods for measuring QA. In a randomized fashion, we prospectively evaluated QA using FD and DD in 48 HD patients during three separate HD sessions over a span of 3 months. The measurement of QA was similar (1,016 +/- 412 ml/min for FD and 1,009 +/- 425 ml/min for DD, p = 0.44 and 0.79 for the mean and standard deviation, respectively). While FD successfully measured QA >or=2,000 ml/min, DD "saturated" (indicating a QA >or=2,000 ml/min without providing a numerical QA value) (n = 17). The correlation coefficient for QA