Automatic blood pressure monitors. Evaluation of three models in volunteers
Luiz Eduardo Imbelloni1, Lúcia Beato, Ana Paula Tolentino
1Instituto de Anestesia Regional, Hospital de Base da FAMERP, Faculdade de Medicina de São José do Rio Preto, SP (FAMERP). imbelloni@openlink.combr
Background And Objectives:
Since 1903, blood pressure has been noninvasively monitored (NIBP), either with manual sphygmomanometer or automated noninvasive devices. One NIBP measurement problem is the considerable variance in blood pressure data, both within and between available techniques. The oscillometric method for NIBP monitoring evaluates blood pressure during cuff deflation. Difficulties in blood pressure measurement by oscillometry may arise from: inadequate cuff size, inadequate cuff application, undetected fails in cuff, hoses, or connectors, arm movement, shock and vascular compression proximal to the cuff. This study aimed at evaluating the reliability of three noninvasive blood pressure monitoring devices during five measurements.
Methods:
Blood pressure of 60 healthy female volunteers aged 20 to 40 years was evaluated from 7 am to 11 am, in the sitting position during a normal workday. Five measures were taken with each device at 2-minute intervals. Three automatic blood pressure monitors were studied. No patient was obese, hypertensive or suffering from cardiac disease and cardiac arrhythmia. Indirect measurements were made according to manufacturers' instructions.
Results:
There were no differences in demographics among the three studied groups. Mean intrapersonal variation from one measurement to the other was up to 6.7 mmHg for systolic blood pressure (SBP), 4.9 mmHg for mean blood pressure (MBP) and 3.3 mmHg for diastolic blood pressure (DBP) with 95% confidence interval. The highest difference between measures in the same volunteer was 49 mmHg for SBP, 46 mmHg for MBP and 28 mmHg for DBP.
Conclusions:
This study has shown significant variations in SBP, MBP and DBP and that SBP is the most reliable parameter to check blood pressure changes in volunteers.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...