[Airflow resistance of shortened tracheal tubes.]
Fernando José Gonçalves do Prado1, Joaquim Edson Vieira, Fábio Ely Martins Benseñor
1CET/SBA, HC da FMUSP.
Background And Objectives:
Tracheal tube length may be shortened just after the cuff probably without noxious influence on airflow resistance. This study aimed at determining the effects of such shortening under different inspiratory flows.
Methods:
Flow resistance was measured in tubes with internal diameters of 7; 7.5; 8; 8.5; 9 and 9.5 millimeters. Measurements were undertaken on standard tubes and on those shortened just after the cuff level. Flows were 0.07 liters per second (L.s-1), 0.1; 0.2; 0.33; 0.5 and 1 L.s-1.
Results:
Resistances were progressively lower for larger internal diameters, but were higher for a same diameter under higher flow, both, in standard and shortened tubes. Shortened tubes showed lower or equal flow resistance as compared to standard tubes with the same diameter.
Conclusions:
Resistance has been lower or comparable on shortened tubes as compared to standard tubes.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Anatomy of Respiratory System II: Lower Respiratory Tract
The Larynx
It is located between the pharynx and the trachea, acts as a passageway for air, and hosts several critical structures, such as the epiglottis, vocal cords, and glottis. The epiglottis acts as a gateway, guiding food to the...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
External and Internal Respiration