Related Experiment Videos
Clinical impact of leak compensation during non-invasive ventilation
Jan Hendrik Storre1, Philipp Bohm, Michael Dreher
1Department of Pneumology, University Hospital Freiburg, Killianstrasse 5, D-79106 Freiburg, Germany.
Background:
This study aimed to assess the impact of leak compensation capabilities during pressure- and volume-limited non-invasive positive-pressure ventilation (NPPV) in COPD patients.
Methods:
Fourteen patients with stable hypercapnic COPD who were receiving long-term NPPV were included in the study. For both modes of NPPV, a full face mask and an artificial leak in the ventilatory circuit were used at three different settings, and applied during daytime NPPV, either without leakage (setting I), with leakage during inspiration only (setting II), and with leakage during inspiration and expiration (setting III). Ventilation pattern was pneumotachy-graphically recorded.
Results:
NPPV was feasible with negligible leak volumes, indicating optimal mask fitting during the daytime (setting I). In the presence of leakage (settings II and III), the attempt to compensate for leak was only evident during pressure-limited NPPV, since inspiratory volumes delivered by the ventilator increased from 726+/-129 (setting I) to 1104+/-164 (setting II), and to 1257+/-166 (setting III) ml during pressure-limited NPPV, respectively (all p<0.001); however, they remained stable during volume-limited NPPV. Leak compensation resulted in a decrease in leakage-induced dyspnea. However, 83%/87% (setting II/III) of the additionally-delivered inspiratory volume during pressure-limited NPPV was also lost via leakage. Expiratory volume was higher in setting II compared to setting III (both p<0.001), indicating the presence of significant expiratory leakage.
Conclusions:
The attempt at leak compensation largely feeds the leakage itself and only results in a marginal increase of tidal volume. However, pressure-limited--but not volume-limited--NPPV results in a clinically-important leak compensation in vivo.
Trial Registration:
www.uniklinik-freiburg.de/zks/live/uklregister/Oeffentlich.html Identifier: UKF001272.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Atelectasis II: Pathophysiology
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management