A new pressurizable dilatometer for measuring the time-dependent bulk modulus and pressure-volume-temperature
Yan Meng1, Paul Bernazzani, Paul A O'Connell
1Department of Chemical Engineering, Texas Tech University, Lubbock, Texas 79409-3121, USA.
Abstract:
A new piston-cylinder-type pressurizable dilatometer controlled by a stepper motor has been developed to measure the time-dependent bulk modulus and pressure-volume-temperature (PVT) behavior of polymeric materials. The dilatometer can be operated from 35 to 230 degrees C and at pressures of up to 250 MPa. The sample cell, which contains the sample and a fluorinated oil as the confining fluid, is totally submerged into a high precision oil bath to achieve a temperature stability of better than 0.01 degrees C. The instrument is calibrated with mercury and quartz. The total instrument volume is 4.0 cm(3), of which 2.3 cm(3) is the sample cell; the total volume can be measured with an average absolute error of better than 5.0x10(-4) cm(3). To demonstrate the instrument's capabilities, the time-dependent bulk modulus and the PVT behavior of a polystyrene are obtained and compared to the literature.