Related Experiment Video
Updated: Jun 22, 2026

Measurement of Outgassing Rates of Steels
Published on: December 13, 2016
Temperature dependence of F-, D- and z-values used in steam sterilization processes
J P C M van Doornmalen1, K Kopinga
13M, Medical Markets Laboratory EMEA, Neuss, Germany.
Aims:
To develop a model, based on microbiological principles, to safely optimize steam sterilization processes.
Methods And Results:
The minimum exposure time F for a decontamination process at a certain temperature is usually calculated from an empirical model with the decimal reduction time D and the temperature resistance coefficient z as parameters. These are implicitly assumed to be independent of temperature. Using a microbiological approach, it is shown that also D and z depend on temperature, indicating that the usual models provide only reliable results in a limited temperature region. The temperature dependence of F resulting from this approach describes the available experimental data very well. Safety margins to assure sterility can be included in a straightforward way.
Conclusions:
The results from the present approach can be used to safely optimize decontamination processes. The corresponding mathematical model can be implemented rather directly in process control systems.
Significance And Impact Of The Study:
Our results show that for steam sterilization and disinfection processes the values of F predicted by the usual models largely underestimate the required minimum exposure times at temperatures below 120 degrees C. This has important consequences for the optimization of such processes.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Temperature
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Thermometers and Temperature Scales
As many physical properties depend on temperature, the variety of thermometers is...
Temperature Dependent Deformation
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...