[Measuring breathing pattern in patients with chronic obstructive pulmonary disease by electrical impedance
Marco Balleza1, Núria Calaf, Teresa Feixas
1Departament de Pneumologia, Hospital de la Santa Creu i de Sant Pau, Facultat de Medicina, Universitat Autònoma de Barcelona, Barcelona, España. jballeza@santpau.cat
Background And Objective:
The measurement of breathing pattern in patients with chronic obstructive pulmonary disease (COPD) by electrical impedance tomography (EIT) requires the use of a mathematical calibration model incorporating not only anthropometric characteristics (previously evaluated in healthy individuals) but probably functional alterations associated with COPD as well. The aim of this study was to analyze the association between EIT measurements and spirometry parameters, static lung volumes, and carbon monoxide diffusing capacity (DLCO) in a group of male patients to develop a calibration equation for converting EIT signals into volume signals.
Materials And Methods:
We measured forced vital capacity (FVC), forced expiratory volume in 1 second (FEV(1)), FEV(1)/FVC, residual volume, total lung capacity, DLCO, carbon monoxide transfer coefficient (KCO) and standard anthropometric parameters in 28 patients with a FEV(1)/FVC ratio of <70%. We then compared tidal volume measurements from a previously validated EIT unit and a standard pneumotachometer.
Results:
The mean (SD) lung function results were FVC, 72 (16%); FEV(1), 43% (14%); FEV(1)/FVC, 42% (9%); residual volume, 161% (44%); total lung capacity, 112% (17%); DLCO, 58% (17%); and KCO, 75% (25%). Mean (SD) tidal volumes measured by the pneumotachometer and the EIT unit were 0.697 (0.181)L and 0.515 (0.223)L, respectively (P<.001). Significant associations were found between EIT measurements and CO transfer parameters. The mathematical model developed to adjust for the differences between the 2 measurements (R(2)=0.568; P<.001) was compensation factor=1.81# - 0.82# height (m)# -0.004 x KCO (%).
Conclusions:
The measurement of breathing pattern by EIT in patients with COPD requires the use of a previously calculated calibration equation that incorporates not only individual anthropometric characteristics but gas exchange parameters as well.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...