Accuracy of whole-body plethysmography requires biological calibration
Porntiva Poorisrisak1, Carsten Vrang1, Jorn Molgaard Henriksen2
1Pediatric Department, Naestved Hospital, Naestved, Denmark.
Background:
Specific airway resistance (sRaw) measured by whole-body plethysmography in young children is increasingly used in research and clinical practice. The method is precise and feasible. However, there is no available method for calibration of the resistance measure, which raises concern of accuracy. Our aim was to determine the agreement of sRaw measurements in six centers and expand normative sRaw values for nonasthmatic children including these centers.
Method:
Identical hardware with different software versions was used at the six centers. Measurements followed a standard operating procedure: (1) seven healthy young children were brought to each of the six centers for sRaw measurements; and (2) 105 healthy preschool children (52 boys; mean age, 5.1 years; interquartile range, 4.3 to 6.0) were recruited locally for sRaw measurements.
Results:
(1) The sRaw of the seven-children study group was significantly lower at two centers compared with the other four centers, and one center had significantly higher sRaw than all the other centers (p < 0.05). Error in the factory settings of the software was subsequently discovered in one of the deviating centers. (2) Normative data (105 preschool children) were generated and were without significant difference between centers and independent of height, weight, age, and gender. We subsequently pooled these normative data (105 children) with our previous data from 121 healthy young children (overall mean sRaw, 1.27; SD, 0.25).
Conclusion:
Control using biological standards revealed errors in the factory setting and highlights the need for developing methods for verification of resistance measures to assure accuracy. Normative data were subsequently generated. Importantly, other centers using such normative data should first consider proper calibration before applying reference values.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.