Evaluation of MVCT protocols for brain and head and neck tumor patients treated with helical tomotherapy
Aude Vaandering1, John Aldo Lee, Laurette Renard
1Department of Radiation Oncology, Université Catholique de Louvain, St-Luc University Hospital, Brussels, Belgium.
Purpose:
Helical tomotherapy is a modality of radiation treatment delivery which is equipped with an on-board imaging device (MVCT) allowing for daily patient set-up verification and correction in the medial-lateral (m-l), cranial-caudal (c-c), anterior-posterior (a-p) and transversal angular (roll) directions. In this study, we measured set-up deviations and evaluated different MVCT protocols for brain and head and neck (H&N) cancer patients.
Materials And Methods:
The daily set-up errors of 75 H&N cancer patients immobilized with 5-point fixation thermoplastic masks and 30 brain cancer patients immobilized with 3-point fixation thermoplastic masks were detected by matching the MVCT with the treatment planning CT images. This co-registration procedure was accomplished automatically by the system's software (automatic deviations), then corrected manually by the radiation therapists (total deviations). Systematic and random errors were analyzed on a patient and a population basis. Moreover, 2 MVCT protocols were retrospectively evaluated; MVCTs were either acquired during the first five fractions (FFFs) or on alternate week (ALT). Systematic deviations were calculated based upon prior "MVCT" fractions and applied during the "non-MVCT" fractions. The resulting residual deviations were then analyzed.
Results:
The total systematic (and random) deviations reached 1.7mm (1.4mm), 1.6mm (1.5mm), 1.5mm (1.5mm) and 0.6 degrees (0.6 degrees ) for H&N cancer patients and reached 1.6mm (0.9mm), 1.7mm (1.1mm), 1.1mm (0.8mm) and 0.9 degrees (0.6 degrees ) for brain cancer patients in the m-l, c-c, a-p and roll directions, respectively. A t-test detected small but statistically significant differences between the automatic and total deviations. Both MVCT protocols gave rise to similar residual deviations. However, for H&N cancer patients the ALT protocol resulted in smaller residual deviations and CTV-PTV margins, particularly in the a-p direction.
Conclusion:
The total systematic and random deviations were comparable to the previously published data. No clinical difference exists between the automatic and total deviations. Both MVCT protocols were similar. But, for H&N cancer patients, the ALT protocol gave rise to smaller residual deviations and therefore is the correct formula to adopt in order to reduce the frequency of pre-treatment MVCTs.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...