Related Experiment Video
Updated: Jun 22, 2026

Fabricating a UV-Vis and Raman Spectroscopy Immunoassay Platform
Published on: November 10, 2016
Competitive homogeneous digoxigenin immunoassay based on fluorescence quenching by gold nanoparticles
Sergiy Mayilo1, Benjamin Ehlers, Michael Wunderlich
1Photonics and Optoelectronics Group, Department of Physics and CeNS, Ludwig-Maximilians-Universität München, Amalienstrasse 54, 80799 Munich, Germany.
Abstract:
We report on a competitive, homogeneous immunoassay for the detection of the hapten digoxigenin. The assay is based on competitive fluorescence quenching by gold nanoparticles. Digoxigenin is indirectly labeled with the fluorophore Cy3B through bovine serum albumin and used as a marker. Gold nanoparticles functionalized with anti-digoxigenin antibodies serve as fluorescence quenchers. Free digoxigenin molecules in the analyte solution compete with the labeled markers for antibodies on the gold nanoparticles. The fluorescence signal depends linearly on the free digoxigenin concentration within a range of concentration from 0.5 to 3 ng mL(-1). The limit of detection is estimated as 0.2 ng mL(-1) and the limit of quantitation is estimated as 0.6 ng mL(-1). The method can be used to detect digoxin, a drug used to cure cardiac arrhythmia.
Related Concept Videos
Immunogold Electron Microscopy
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.