Related Experiment Video
Updated: Jun 22, 2026

Fluorescence-quenching of a Liposomal-encapsulated Near-infrared Fluorophore as a Tool for In Vivo Optical Imaging
Published on: January 5, 2015
A primary method for determination of optical parameters of turbid samples and application to intralipid between 550
Cheng Chen1, Jun Q Lu, Huafeng Ding
1Department of Physics, East Carolina University, Greenville, NC 27858, USA.
Abstract:
The lack of a primary method for determination of optical parameters remains a significant barrier in optical study of turbid media. We present a complete system of experimental setups and Monte Carlo modeling tools for fast and accurate solution of the inverse problem from the measured signals of homogeneous turbid samples. The calibration of the instrument and validation of the Monte Carlo modeling have been carried out to ensure the accuracy of the inverse solution. We applied this method to determine the optical parameters of turbid media of 10% intralipid between 550 and 940 nm and 20% intralipid between 550 and 1630 nm.
Related Concept Videos
Total Internal Reflection Fluorescence Microscopy
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
UV–Vis Spectrometers