Can mixed venous hemoglobin oxygen saturation be estimated using a NICO monitor?
Yoshifumi Kotake1, Takashige Yamada, Hiromasa Nagata
1Department of Anesthesiology, Toho University Medical Center Ohmori Hospital, 6-11-1, Ohmori-nishi, Ohta, Tokyo 143-8541, Japan. ykotake@med.toho-u.ac.jp
Background:
We hypothesized that mixed venous hemoglobin oxygen saturation (SvO2) can be estimated by calculation from CO2 production, cardiac output, and arterial oxygen saturation measured using a noninvasive cardiac output (NICO) monitor (Novametrix-Respironics, Wallingford, CT).
Methods:
Twenty-three patients undergoing aortic aneurysm repair underwent SvO2 monitoring using a pulmonary artery catheter and cardiac output monitoring using a NICO monitor. The estimated SvO2 value calculated from NICO monitor-derived values was compared with the SvO2 value measured using a pulmonary artery catheter. The accuracy of this estimation was analyzed with Bland-Altman method. The ability of this estimation to track the change of SvO2 was also evaluated using correlation analysis to compare the changes of estimated SvO2 and measured SvO2.
Results:
The bias +/- limits of agreement of the estimated SvO2 against measured SvO2 was -2.1% +/- 11.2%. The change of estimated SvO2 was modestly correlated with the change of measured SvO2.
Conclusions:
SvO2 derived from the values measured by the NICO monitor cannot be used interchangeably with the values measured spectrophotometrically using the pulmonary artery catheter. More refinement is required to obtain more reliable estimate of SvO2 less invasively. However, large changes of SvO2 may be detected with this method and can be used as a precautionary sign when the balance between oxygen supply and demand is compromised without inserting a central venous catheter.
More Related Videos
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
08:23Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Transport in the Blood
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...