External fixator clamp reuse degrades clamp mechanical performance
Robert S Gilley1, David P Beason, David M Snyder
1Department of Clinical Studies, School of Veterinary Medicine, University of Pennsylvania, Philadelphia, PA 19104, USA. rsgilley@vet.upenn.edu
Reusable external fixator clamps degrade with repeated use. The IMEX-SK and Securos clamps showed more significant mechanical property loss than the Kirschner-Ehmer (KE) clamp, impacting fracture stability.
Area of Science:
- Orthopedic biomechanics
- Biomaterials science
- Surgical instrumentation
Background:
- External fixator clamps are crucial for maintaining fracture reduction stability.
- The mechanical integrity of these clamps is essential for successful fracture healing.
- The impact of clamp reuse on mechanical properties is not fully understood.
Purpose of the Study:
- To evaluate the effects of clamp reuse on the mechanical properties of three common external fixator clamp types: Kirschner-Ehmer (KE), Securos, and IMEX-SK.
- To compare the degradation patterns among the tested clamp types.
Main Methods:
- An experimental bench test was conducted to assess the mechanical properties of 18 clamps of each type (KE, Securos, IMEX-SK).
- Six mechanical variables were tested: fixator pin slippage, connecting bar slippage, fixator pin rotation, connecting bar rotation, and clamp-bolt axis pivot (clockwise and counterclockwise).
- Each clamp underwent six repeated tests for each variable at a consistent clamp-bolt tightening torque of 7.68 Nm.
Main Results:
- All three clamp types exhibited degradation in at least one mechanical variable after reuse.
- The IMEX-SK clamp showed significant degradation in connecting bar slippage and fixator pin rotation, with an increased resistance to clockwise clamp-bolt axis pivot.
- The Securos clamp degraded significantly in connecting bar slippage, connecting bar rotation, and fixator pin rotation, while the KE clamp showed degradation only in connecting bar slippage.
Conclusions:
- Repeated use of external fixator clamps leads to degradation of their mechanical performance, affecting their ability to resist motion.
- The IMEX-SK and Securos clamps demonstrated greater susceptibility to degradation compared to the KE clamp.
- Clinicians should be aware of the potential for reduced fixator composite rigidity and unpredictable performance when reusing external fixator clamps.
More Related Videos
09:26An Efficient and Reproducible Protocol for Distraction Osteogenesis in a Rat Model Leading to a Functional Regenerated Femur
Published on: October 23, 2017
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Residual Stresses
Temperature Dependent Deformation