External fixator clamp reuse degrades clamp mechanical performance
Robert S Gilley1, David P Beason, David M Snyder
1Department of Clinical Studies, School of Veterinary Medicine, University of Pennsylvania, Philadelphia, PA 19104, USA. rsgilley@vet.upenn.edu
Objective:
To determine the effects of clamp reuse for the Kirschner-Ehmer (KE); Securos; and the IMEX-SK clamp.
Study Design:
Experimental bench test of mechanical properties.
Methods:
Specially designed fixtures were used to mechanically test 18 clamps of each type with respect to 6 mechanical variables: fixator pin slippage, connecting bar slippage, fixator pin rotation, connecting bar rotation, and clockwise and counterclockwise clamp-bolt axis pivot. Each clamp was tested 6 times for each variable at 7.68 Nm of clamp-bolt tightening torque.
Results:
were compared using repeated measures ANOVA. Results- For the IMEX-SK clamp, a significant degradation in the force required to cause slippage was found for connecting bar slippage and fixator pin rotation; however, this clamp also had a significant increase in the force to initiate slippage for clockwise clamp-bolt axis pivot with reuse. The Securos clamp had significant degradation in connecting bar slippage, connecting bar rotation, and fixator pin rotation whereas the KE clamp had significant degradation in connecting bar slippage only.
Conclusions:
All 3 external fixator clamp types degraded in 1 or more movement variables in their ability to resist motion with reuse. The IMEX and Securos clamps were more subject to degradation than the KE clamp and this may have clinical importance for fixator composite rigidity.
Clinical Relevance:
Fracture-reduction stability is related to the ability to resist motion within a clamp. The unpredictable nature of degradation we found cautions against repeated use. Clinicians should consider reuse of external fixator clamps with the knowledge that repeated use degrades clamp mechanical performance.
More Related Videos
09:26An Efficient and Reproducible Protocol for Distraction Osteogenesis in a Rat Model Leading to a Functional Regenerated Femur
Published on: October 23, 2017
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Residual Stresses
Temperature Dependent Deformation