Minimum refractometrically detectable concentrations of single atmospheric gases and simple mixtures with a Sagnac
1Applied Optics Program, Queensland University of Technology, 2 George Street, Brisbane, Queensland, 4000, Australia. s.mcconnell@qut.edu.au
Abstract:
The minimum quantities of the nine most abundant, isolated, atmospheric gases that are detectable with a refractometer are calculated. An examination of the applicability of refractometric techniques for detecting and analyzing gaseous mixtures is discussed and a comparison made against other established techniques. Traditionally, most gas analysis performed with an interferometer is in determining the dispersion or refractivity of a known sample, presented here is the inverse approach, where refractivities are measured to determine the concentrations of particular species within a gas. The method, and experimental results for determining the minimum quantities of a particular species detectable in a mixture has been explored, as well as the complications, such as the indistinguishability of dynamic polarizabilities of different gases and the subsequent demands for accurate pressure and fringe measurements of using interferometric techniques. It is shown that the concentration of a single (isolated) gas, in units of number density, can be determined to within approximately 1-10 x 10(18) m(-3), and a mixture of the three most abundant gases, N2, O2 and Ar, to within 3.4 x 10(4) parts in 10(6) (ppm) when a minimum detectable fringe shift of lambda/100 is assumed.
Related Concept Videos
IR Spectrometers
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Gas Chromatography: Types of Detectors-II
UV–Vis Spectrometers
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...