Is humeral segmental defect replacement device a stronger construct than locked IM nailing?
Robert Heck1, Ruxandra Marinescu, Haden Janda
1InMotion Musculoskeletal Institute, Memphis, TN 38103, USA. rheck@campbellclinic.com
Clinical Orthopaedics and Related Research
|June 23, 2009
Summary
Humeral segmental replacement prostheses offer a stronger construct than intramedullary (IM) nails for impending pathologic fractures. This improved strength is particularly evident in weaker bones, suggesting a better treatment option.
Area of Science:
- Orthopedic surgery
- Biomechanical engineering
- Bone pathology
Background:
- Intramedullary (IM) nailing is standard for impending pathologic humeral fractures.
- Rotator cuff violation during IM nail insertion causes complications.
- A superior fixation method is needed for these fractures.
Purpose of the Study:
- To compare the mechanical properties of humeral segmental replacement prostheses and IM nails.
- To evaluate construct strength in a simulated impending pathologic humeral fracture model.
Main Methods:
- Nine matched pairs of human humeri were used.
- A 50% middiaphyseal defect simulated an impending fracture.
- Specimens were fixed with either an IM nail or a segmental defect prosthesis.
- Mechanical testing in torsion to failure was performed.
Main Results:
- Prosthesis constructs demonstrated greater peak torque and peak rotation at failure.
- Intramedullary nail constructs exhibited higher torsional stiffness.
- A linear relationship between peak torque and bone T-score was observed for both devices.
- The prosthesis showed greater superiority in weaker bones (lower T-scores).
Conclusions:
- Humeral segmental replacement prostheses provide a stronger construct than IM nails for impending pathologic humeral fractures.
- The prosthesis is a potentially advantageous alternative, especially in osteoporotic bone.