Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Systematic uncertainties in the Monte Carlo calculation of ion chamber replacement correction factors
L L W Wang1, D J La Russa, D W O Rogers
1Ottawa Carleton Institute of Physics, Carleton University, Campus Ottawa, Ottawa, Ontario KIS 5B6, Canada. lwang@physics.carleton.ca
Abstract:
In a previous study [Med. Phys. 35, 1747-1755 (2008)], the authors proposed two direct methods of calculating the replacement correction factors (P(repl) or P(cav)P(dis)) for ion chambers by Monte Carlo calculation. By "direct" we meant the stopping-power ratio evaluation is not necessary. The two methods were named as the high-density air (HDA) and low-density water (LDW) methods. Although the accuracy of these methods was briefly discussed, it turns out that the assumption made regarding the dose in an HDA slab as a function of slab thickness is not correct. This issue is reinvestigated in the current study, and the accuracy of the LDW method applied to ion chambers in a 60Co photon beam is also studied. It is found that the two direct methods are in fact not completely independent of the stopping-power ratio of the two materials involved. There is an implicit dependence of the calculated P(repl) values upon the stopping-power ratio evaluation through the choice of an appropriate energy cutoff delta, which characterizes a cavity size in the Spencer-Attix cavity theory. Since the delta value is not accurately defined in the theory, this dependence on the stopping-power ratio results in a systematic uncertainty on the calculated P(repl) values. For phantom materials of similar effective atomic number to air, such as water and graphite, this systematic uncertainty is at most 0.2% for most commonly used chambers for either electron or photon beams. This uncertainty level is good enough for current ion chamber dosimetry, and the merits of the two direct methods of calculating P(repl) values are maintained, i.e., there is no need to do a separate stopping-power ratio calculation. For high-Z materials, the inherent uncertainty would make it practically impossible to calculate reliable P(repl) values using the two direct methods.
Related Concept Videos
Propagation of Uncertainty from Systematic Error
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Uncertainty in Measurement: Accuracy and Precision
Random and Systematic Errors
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...