Wavy horizontal artifacts on optical coherence tomography line-scanning images caused by diffractive multifocal
Makoto Inoue1, Hiroko Bissen-Miyajima, Mami Yoshino
1Department of Ophthalmology, Tokyo Dental College Suidobashi Hospital, Kyorin University School of Medicine, Tokyo, Japan. inoue@eye-center.org
Purpose:
To report wavy horizontal artifacts on optical coherence tomography (OCT) monitor images.
Setting:
Tokyo Dental College Suidobashi Hospital, Department of Ophthalmology, Tokyo, Japan.
Methods:
The line-scanning ophthalmoscopy images from an OCT device of eyes that had uneventful cataract extraction with implantation of a diffractive multifocal intraocular lens (IOL) were evaluated. The images were compared with those of eyes that had cataract surgery with implantation of a monofocal IOL.
Results:
The study evaluated 38 eyes of 19 patients with a diffractive multifocal IOL and 29 eyes of 18 patients with a monofocal IOL. Wavy horizontal artifacts were seen in all eyes with a multifocal IOL but not in any image of eyes with a monofocal IOL (P<.0001, Fisher exact probability test). The OCT images, fundoscopic photographs, and scanning-laser ophthalmoscopy images were unaffected by the multifocal IOLs.
Conclusion:
The aberrations in the images from line-scanning ophthalmoscopy may have been caused by the optical design of the diffractive multifocal IOLs.
More Related Videos
07:44In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
06:55Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Focusing of Light in the Eye
Interference and Diffraction
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...