64-Detector CT angiography in renal artery stent evaluation: prospective comparison with selective catheter
Clemens Steinwender1, Wilhelm Schützenberger, Franz Fellner
1Cardiovascular Division, Linz City Hospital, Krankenhausstrasse 9, A-4020 Linz, Austria. clemens.steinwender@akh.linz.at
Purpose:
To prospectively assess the diagnostic accuracy of 64-detector computed tomographic (CT) renal artery (RA) angiography for the evaluation of RA in-stent restenosis (ISR) by using selective catheter RA angiography as the reference standard.
Materials And Methods:
Institutional review board approval and written informed consent were obtained. Eighty-six patients (46 men, 40 women; mean age +/- standard deviation, 71 years +/- 9) with 95 RA stents were examined with 64-detector CT RA angiography by experienced radiologists 1 day before selective catheter RA angiography was performed by experienced cardiologists. Each group was blinded to the results of the other imaging method. For image reconstruction, a standard (B25f) and an edge-enhancing (B46f) kernel were chosen. An ISR of more than 50% was considered as hemodynamically significant. Sensitivity, specificity, and positive and negative predictive values for the detection of ISR by using CT RA angiography were calculated.
Results:
At CT RA angiography, 93 (98%) of 95 stents were assessable for diagnosis. Two stents could not be evaluated owing to hardening artifacts affected by vessel calcifications. All nine cases of significant ISR seen at selective catheter RA angiography were correctly diagnosed by using CT RA angiography, giving a sensitivity of 100% and a negative predictive value of 100%. One case of nonsignificant ISR seen at selective catheter RA angiography was interpreted as significant by using CT RA angiography, giving a specificity of 99% and a positive predictive value of 90%. In four of 78 patients without ISR seen at selective catheter RA angiography, CT RA angiography showed nonsignificant ISR, giving a specificity of 95% and a positive predictive value of 56%.
Conclusion:
Sixty-four-detector CT RA angiography can provide an excellent noninvasive technique to help detect and evaluate ISR within the RA stents used in our study.
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...