A Novel Technique for Endovascular Detection and Removal of Radiographic Contrast during Angiography
Hyeonsoo Chang1, Ali H M Hassan, Young L Kim
1Center for Cardiovascular Technology, Stanford University Medical Center, CA 94305, USA.
Objectives:
This study aims at in-vitro validation of the principles of endovascular detection of contrast medium and assessing the feasibility of in-vivo detection and removal of contrast during angiography.
Background:
Contrast-induced nephropathy is a growing concern in current percutaneous interventions with increasing lesion complexity and patient comorbidity. To address this clinical problem, a novel method of endovascular detection and automatic removal of contrast has been developed, and is comprised of a catheter-based system with a reflectance-type optical sensor.
Methods:
Blood samples were obtained from ovine subjects to characterize the optical response of blood by measuring the reflectance spectrum at varying levels of hematocrit diluted by a contrast agent. The results from the in-vitro test were implemented into an in-vivo system. An aspiration catheter equipped with a fiberoptic sensor was inserted into the coronary sinus (CS) of 5 canines. Contrast was administered through the coronary artery and reflectance signals were recorded at the CS. The removal rate was analyzed through 20 specimen collections.
Results:
A proportional relationship was found between hematocrit and reflectance intensity in in-vitro test. Upon in-vivo detection of contrast, the sensor signal showed a 79.5 +/- 9.9% (n = 33) drop from the pre-injection baseline. This was highly reproducible and beyond the noise level of baseline, (2.5 +/- 0.9%), enabling automatic activation of the aspiration system. The signal duration was 12.2 +/- 3.7 seconds. The removal rate of contrast was 59.3 +/- 11%.
Conclusion:
The present study validated the principles of endovascular contrast detection and demonstrated the feasibility of an in-vivo, catheter-based removal of contrast using reflectance technology.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
05:26Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
Published on: January 7, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies III: Computed Tomography