Evaluation of intramedullary rib splints for less-invasive stabilisation of rib fractures
Inga Helzel1, William Long, Daniel Fitzpatrick
1Biomechanics Laboratory, Legacy Research & Technology Centre, Portland, OR 97232, USA.
Background:
Intramedullary fixation of rib fractures with generic Kirschner wires has been practiced for over 50 years. However, this technique has not been advanced to address reported complications of wire migration and cut-out. This biomechanical study evaluated a novel rib splint designed to replicate the less-invasive fixation approach of Kirschner wires while mitigating their associated complications.
Methods:
The durability, strength, and failure mode of rib fracture fixation with intramedullary rib splints were evaluated in 27 cadaveric ribs. First, intact ribs were loaded to failure to determine their strength and to induce realistic rib fractures. Subsequently, fractures were stabilised with a novel rib splint made of titanium alloy with a rectangular cross-section that was secured with a locking screw. All fixation constructs were dynamically loaded to 360,000 cycles at five times the respiratory load magnitude to determine their durability. Finally, constructs were loaded to failure to determine their residual strength and failure modes.
Results:
Native ribs had a strength of 9.7+/-5.0 N m, with a range of 3.5-19.6 N m. Fracture fixation with rib splints was uneventful. All 27 splint constructs sustained dynamic loading without fixation failure, implant migration or implant cut-out. Dynamic loading caused no significant decrease in construct stiffness (p=0.85) and construct subsidence remained on average below 0.5 mm. The residual strength of splint constructs after dynamic loading was 1.1+/-0.24 N m. Constructs failed by splint bending in 44% of specimens and by developing fracture lines along the superior and inferior cortices in 56% of specimens. Regardless of the failure mode, all rib splint constructs recoiled elastically after failure and retained functional reduction and fixation. No construct exhibited implant cut-out or migration through the lateral cortex.
Conclusions:
Rib splints can provide sufficient stability to support respiratory loading throughout the healing phase, but they cannot restore the full strength of native ribs. Most importantly, rib splints mitigated the complications reported for rib fracture fixation with generic Kirschner wires, namely implant cut-out and migration through the lateral cortex. Therefore, rib splints may provide an advanced alternative to the original Kirschner wire technique for less-invasive fixation of rib fractures.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...