Related Experiment Video
Updated: Jun 21, 2026

Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
Published on: April 30, 2018
Temperature mapping near the surface of ultrasound transducers using susceptibility- compensated magnetic resonance
Andrew G Webb1, Thomas Neuberger, Eun-Joo Park
1Department of Bioengineering, The Pennsylvania State University, University Park, PA, USA.
Abstract:
Magnetic resonance imaging (MRI)-based temperature mapping very close to the surface of an ultrasound transducer is not possible due to the large magnetic susceptibility- induced image artifacts that arise from the materials used in transducer construction. Here, it is shown in phantoms that "susceptibility-compensated" MRI sequences can be used to measure thermal increases approximately 1 mm from the surface of a 4-element cymbal array transducer, which has been used widely for noninvasive transdermal drug delivery. The estimated temperatures agree well with those measured using thermocouples.
