Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Extending the working calibration ranges of four hexachlorocyclohexane isomers in gas chromatography-electron capture
Yanfei Chen1, Jin He, Jibin Zhang
1State Key Laboratory of Agricultural Microbiology and National Engineering Research Center of Microbial Pesticides, Huazhong Agricultural University, Wuhan 430070, PR China.
Abstract:
A radial basis function neural network (RBFNN) method was developed for the first time to model the nonlinear calibration curves of four hexachlorocyclohexane (HCH) isomers, aiming to extend their working calibration ranges in gas chromatography-electron capture detector (GC-ECD). Other 14 methods, including seven parametric curve fitting methods, two nonparametric curve fitting methods, and five other artificial neural network (ANN) methods, were also developed and compared. Only the RBFNN method, with logarithm-transform and normalization operation on the calibration data, was able to model the nonlinear calibration curves of the four HCH isomers adequately. The RBFNN method accurately predicted the concentrations of HCH isomers within and out of the linear ranges in certified test samples. Furthermore, no significant difference (p>0.05) was found between the results of HCH isomers concentrations in water samples calculated with RBFNN method and ordinary least squares (OLS) method (R(2)>0.9990). Conclusively, the working calibration ranges of the four HCH isomers were extended from 0.08-60 ng/ml to 0.08-1000 ng/ml without sacrificing accuracy and precision by means of RBFNN. The outstanding nonlinear modeling capability of RBFNN, along with its universal applicability to various problems as a "soft" modeling method, should make the method an appealing alternative to traditional modeling methods in the calibration analyses of various systems besides the GC-ECD.
Related Concept Videos
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Resolution Mass Spectrometry (HRMS)