Evaluation of the PAS-Port Proximal Anastomosis System in coronary artery bypass surgery (the EPIC trial)

John D Puskas1, Michael E Halkos, Husam Balkhy

  • 1Division of Cardiothoracic Surgery, Emory University School of Medicine, Atlanta, Georgia 30308, USA. john.puskas@emoryhealthcare.org

Insights

The PAS-Port device offers comparable graft patency to traditional hand-sewn anastomoses in coronary artery bypass grafting. This automated device reduces surgical time without compromising patient outcomes.

Area of Science:

  • Cardiovascular Surgery
  • Medical Devices
  • Surgical Innovation

Background:

  • Traditional coronary artery bypass grafting (CABG) often requires aortic clamping for proximal vein graft anastomosis.
  • This clamping can be associated with potential complications and increased surgical time.

Purpose of the Study:

  • To evaluate the efficacy of the PAS-Port device for automated proximal anastomosis in CABG.
  • To compare the PAS-Port device to traditional hand-sewn anastomoses without aortic clamping.

Main Methods:

  • A multicenter, prospective, randomized trial involving 220 patients undergoing CABG.
  • Randomized assignment of either PAS-Port device or hand-sewn anastomosis for proximal vein grafts within each patient.
  • Primary endpoint: angiographic graft patency at 9 months; Secondary endpoints: anastomotic time and major adverse cardiac events.

Main Results:

  • 9-month graft patency was 80.3% for PAS-Port versus 82.0% for hand-sewn, demonstrating non-inferiority.
  • Freedom from major adverse cardiac events at 9 months was high for both groups (97.7% PAS-Port, 98.2% hand-sewn).
  • The PAS-Port device significantly reduced anastomotic time by an average of 4.6 minutes (P < .001).

Conclusions:

  • The PAS-Port device provides an effective proximal anastomosis in CABG.
  • It achieves comparable 9-month graft patency to hand-sewn anastomoses.
  • The device enables anastomosis without aortic clamping, reducing surgical time.