Related Experiment Video
Updated: Jun 21, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Change in glucometer settings as a cause of sudden deterioration of glycemic control in type 2 diabetes
Zdenek Jankovec1, Silvie Lacigova
1Department of Medicine I, Charles University Hospital, Plzen, Czech Republic. jankovec@fnplzen.cz
Abstract:
The subject matter of this case report is "sudden deterioration" of glycemic control in a thus far well-complying patient with type 2 diabetes. It describes possible impacts of glucometer technical failure associated with other unfavorable circumstances. An error in displaying glycemia was discovered when analyzing data from glucometer in a computer using software (DIABASS PRO, Mediaspect GmbH, Konstanz, Germany) for data evaluation, and other possible complications (especially hypoglycemia episodes) resulting from inadequate treatment correction were thus prevented.