Related Experiment Video
Updated: Jun 21, 2026

Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
Use of a quality-by-design approach to justify removal of the HPLC weight % assay from routine API stability testing
Peter J Skrdla1, Tao Wang, Vincent Antonucci
1Merck & Co., Inc., P.O. Box 2000, 126 East Lincoln Avenue, RY818-B205, Rahway, NJ 07065, USA. peter_skrdla@merck.com
Abstract:
Due to the high method variability (typically > or = 0.5%, based on a literature survey and internal Merck experience) encountered in the HPLC weight percent (%) assays of various active pharmaceutical ingredients (APIs), it is proposed that the routine use of the test in stability studies should be discouraged on the basis that it is frequently not sufficiently precise to yield results that are stability-indicating. The high method variability of HPLC weight % methods is not consistent with the current ICH practice of reporting impurities/degradation products down to the 0.05% level, and it can lead to erroneous out-of-specification (OOS) results that are due to experimental error and are not attributable to API degradation. For the vast majority of cases, the HPLC impurity profile provides much better (earlier and more sensitive) detection of low-level degradation products. Based on these observations, a Quality-by-Design (QbD) approach is proposed to phase out the HPLC weight % assay from routine API stability testing protocols.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Clinically Relevant Drug Product Specifications: Methods of Establishment
In Vitro Drug Release Testing: Overview, Development and Validation
Data Validation
Key parameters for method validation include:
High-Performance Liquid Chromatography: Elution Process