Development of inlaid electrodes for whole column electrochemical detection in HPLC
Jung-Ho Seo1, Pei Ling Leow, Si-Hyeong Cho
1Micro Biochip Center, Hanyang University, Ansan, 426-791, Korea.
Abstract:
An electrochemical microfluidic device has been fabricated on PET (polyethylene terephthalate) substrate using an imprinting method. The imprinting transfers patterns from a stamp into a substrate mechanically. However, a blanket mould imprinting process has been introduced to embed the photolithographically produced gold metal electrode lines into the PET substrate resulting in an individually addressable array flush to better than 100 nm. The device formed one wall of a packed chromatography column. The array was electrochemically characterised using standard redox probes in both stagnant conditions and under flow. Both numerical modelling and experimental data show improved sensitivity under flow and a limiting current which scaled linearly with the cube root of the volume flow rate. A chromatographic separation of the bioanalytical significant neurotransmitter dopamine (DA) and its metabolite DOPAC was achieved and electrochemically detected at multiple locations within the column. The PET device was stable and robust to leaks to pressures well in excess of those required for chromatographic separations.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Electrodeposition
Electrodeposition can...