Background:
Although inspiratory assist of spontaneous breathing in intubated patients is common, expiratory assist functions have rarely been reported. Effective expiratory support (ES) could be used to compensate for endotracheal tube (ETT) resistance during spontaneous breathing. In this study, we examined the performance of a new system designed to provide both inspiratory support (IS) and ES during spontaneous breathing with the goal of reducing the effective resistance of the ETT.
Methods:
The ES system consisted of a ventilator demand valve and a computer-controlled piston cylinder, which aspirated gas from the respiratory circuit during the expiratory phase. The movement of the piston was synchronized with spontaneous breathing. We compared the pressures at the tip of the ETT and in the breathing circuit during spontaneous breathing through an ETT of internal diameter (ID) 5 mm with that of an ETT with ID 8 mm in nine healthy adult male volunteers. The ventilatory mode was set to maintain a continuous airway pressure of 0 cm H(2)O. Three ventilator settings (no support, IS only, and IS plus ES) were compared using ID 5 mm ETT.
Results:
We monitored pressure in the breathing circuit (P(aw)), ETT tip pressure (P(tip)), and respiratory flow. The P(tip) of the ID 5 mm ETT showed a large negative deflection during inspiration and a positive deflection during expiration without support. IS alone did not improve the respiratory pattern through the small ETT. However, IS plus ES resulted in negative P(aw) during expiration in addition to positive deflection of P(aw) during inspiration, making the pressure characteristics of P(tip) similar to those of ID 8 mm ETT. Moreover, IS plus ES produced a respiratory pattern through the ID 5 mm ETT that was similar to that through the ID 8 mm ETT.
Conclusions:
In this study of healthy volunteers, IS plus ES compensated for the airway resistance imposed by a ID 5.0 mm ETT to create pressure changes at the tip of the ETT similar to those of an ID 8.0 mm ETT.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pulmonary Cycle: Exhalation
Mechanism of Breathing III: The Accessory Muscles
The respiratory system is a complex network that relies on primary respiratory muscles like the diaphragm, but also involves accessory muscles to enhance lung expansion and airflow during both inhalation and exhalation.
Enhancing Inhalation with Accessory Muscles:
Accessory muscles such as the sternocleidomastoid, scalene, intercostal, and abdominal muscles are crucial when additional respiratory effort is required, such as during deep...