Objective:
The aim of this study was to evaluate the mechanical stresses over the bone and hardware after sagittal split ramus osteotomy (SSRO) fixed with standard titanium or locking plate/screws using finite element analysis.
Study Design:
A 3-dimensional finite element model of the mandible was created, and SSRO and 5 mm advancement was simulated on a computer model. The model was fixed with either 2.0-mm titanium conventional miniplate/screw or 2.0-mm titanium locking miniplate/screw system, and oblique 200 N bite force was applied.
Results:
The values of von Mises stresses in the cortical layer of the distal segment using the locking plate system was higher. However, in the cortical layer of the proximal segment the stresses were higher at conventional plate system. In the spongiosa layers of both segments, stresses were higher with the conventional plate system.
Conclusion:
The locking miniplate/screw system spreads the load over the plate and screws and diminishes the amount of force transfered to each unit.
More Related Videos
07:16Finite Element Analysis Model for Assessing Expansion Patterns from Surgically Assisted Rapid Palatal Expansion
Published on: October 20, 2023
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...
Stability of structures