Evaluation of radiation doses from MDCT-imaging in otolaryngology
Chiyo Yamauchi-Kawaura1, Keisuke Fujii, Takahiko Aoyama
1School of Health Sciences, Nagoya University, 1-1-20, Daikominami, Higashi-ku, Nagoya, Japan. kawaura@met.nagoya-u.ac.jp
Abstract:
The purpose of this study was to clarify patient doses in the current otolaryngological multi-detector row computed tomography (MDCT) examinations. Patient doses were measured with an in-phantom dosimetry system which was composed of 48 photodiode dosimeters embedded within an anthropomorphic phantom. Organ and effective doses were evaluated according to the International Commission on Radiological Protection Publication 103. In neck CT, doses for salivary glands and for thyroid were high, 7.6-29.9 and 13.4-60.3 mGy, respectively. In sinus CT, brain and lens doses were high, 7.6-24.6 and 10.6-32.0 mGy, respectively, and in inner ear CT, lens dose was 8.0-35.3 mGy. Effective doses were 1.8-6.6 mSv in neck CT, 0.5-0.9 mSv in sinus CT and 0.3-0.6 mSv in inner ear CT. The present dose data would be used to estimate radiation risks for patients undergoing otolaryngological MDCT examinations.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...