Rune Thorsen1, Maurizio Ferrarin
1Polo Tecnologico (Biomedical Technology Department), IRCCS S. Maria Nascente, Fondazione Don Carlo Gnocchi Onlus, Via Capecelatro, 66 20148 Milano, Italy. rthorsen@dongnocchi.it
Abstract:
Surface Functional Electrical Stimulation (FES) requires high stimulation voltages. A step-up transformer in the output stage of the stimulation circuit is often used. In the present technical paper a voltage controlled current source (VCCS) is presented as an alternative to the transformer coupling. Two (master-slave) coupled transconductance amplifiers (TAs)--in series with pre-charged capacitors--are used to drive the output current. After each stimulation pulse the capacitors are recharged to a high voltage by a switch mode power supply (SMPS). A multiplexer in the output stage is used to provide biphasic output. Output rise-time (10-90%) was less than 2 micros at 100 mA output. Biphasic charge balanced stimulation current can be produced with a net current to ground of less than 20 nA, thus virtually separated from ground. The circuit permits recording of the volitional myoelectric signal from the stimulated muscle. It is part of a portable myoelectrically controlled FES system powered by 2 AA batteries and currently used in clinical trials.