Using Laser Doppler Imaging and Monitoring to Analyze Spinal Cord Microcirculation in Rat
Published on: May 30, 2018
Blood flow measurements at different depths using photoplethysmography and laser Doppler techniques
Sara Bergstrand1, Lars Göran Lindberg, Anna-Christina Ek
1Department of Medical and Health Sciences, Linköping University, Linköping, Sweden. sara.bergstrand@imv.liu.se
Background/Purpose:
This study has evaluated a multi-parametric system combining laser Doppler flowmetry and photoplethysmography in a single probe for the simultaneous measurement of blood flow at different depths in the tissue. This system will be used to facilitate the understanding of pressure ulcer formation and in the evaluation of pressure ulcer mattresses.
Methods:
The blood flow in the tissue over the sacrum was measured before, during and after loading with 37.5 mmHg, respectively, 50.0 mmHg. The evaluation of the system consisted of one clinical part, and the other part focusing on the technicalities of the probe prototype.
Results:
An increase in blood flow while loading was the most common response, but when the blood flow decreased during loading it was most affected at the skin surface and the blood flow responses may be different due to depths of measurement. Reactive hyperaemia may occur more frequently in the superficial layers of the tissue.
Conclusion:
The study showed that the new system is satisfactory for measuring tissue blood flow at different depths. The laser Doppler complements the photoplethysmography, and further development of the system into a thin flexible probe with the ability to measure a larger area is required.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...