Bacterial contamination of PCA and epidural infusion devices
M Rothwell1, D Pearson, K Wright
1Acute Pain Service, Macclesfield Hospital, Macclesfield, Cheshire, UK. mick@ohc.me.uk
Abstract:
We prospectively audited the bacterial contamination of re-usable analgesia infusion pumps. In a one-month period, 112 samples from the handset and keypads of our analgesia infusion pumps were cultured for bacterial contamination. Forty-five percent of handset swabs and 46% of keypad swabs grew bacteria; the commonest organism being coagulase-negative staphylococcus. An additional cleaning step using 70% isopropyl alcohol wipes was introduced and the contamination rate was re-audited in 100 samples. The contamination rate was reduced to 6% of handset swabs and 4% of keypad swabs. A high initial rate of bacterial contamination of re-usable analgesia infusion pumps was significantly reduced by the implementation of a simple, additional cleaning procedure.
Insights
Bacterial contamination of reusable analgesia infusion pumps was high. An extra cleaning step using isopropyl alcohol wipes significantly reduced pump contamination, improving patient safety.
Area of Science:
- Microbiology
- Infection Control
- Healthcare Technology
Background:
- Reusable medical devices, such as analgesia infusion pumps, can harbor bacteria.
- High bacterial contamination rates pose a risk of healthcare-associated infections.
Purpose of the Study:
- To audit bacterial contamination on reusable analgesia infusion pumps.
- To evaluate the effectiveness of an additional cleaning protocol.
Main Methods:
- Prospective audit of 112 samples from pump handsets and keypads.
- Bacterial cultures to identify contamination.
- Implementation of 70% isopropyl alcohol wipe cleaning.
- Re-audit of 100 samples after implementing the new cleaning protocol.
Main Results:
- Initial audit revealed 45% handset and 46% keypad contamination, primarily with coagulase-negative staphylococcus.
- Post-intervention audit showed significantly reduced contamination rates: 6% for handsets and 4% for keypads.
- The additional cleaning step demonstrated high efficacy in reducing bacterial load.
Conclusions:
- Reusable analgesia infusion pumps are frequently contaminated with bacteria.
- A simple cleaning protocol using 70% isopropyl alcohol wipes is highly effective in reducing contamination.
- Implementing enhanced cleaning procedures is crucial for infection control in healthcare settings.
Related Concept Videos
Endocarditis IV: Nursing Management
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies. Common...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Bacterial Signaling