Related Experiment Video
Updated: Jun 21, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Rigid fibrescope Bonfils: use in simulated difficult airway by novices
Tim Piepho1, Rüdiger R Noppens, Florian Heid
1University Medical Center of the Johannes Gutenberg-University, Department of Anaesthesiology, Mainz, Germany. tim.piepho@gmx.de
Background:
The Bonfils intubation fibrescope is a promising alternative device for securing the airway. We examined the success rate of intubation and the ease of use in standardized simulated difficult airway scenarios by physicians. We compared the Bonfils to a classical laryngoscope with Macintosh blade.
Methods:
30 physicians untrained in the use of rigid fibrescopes but experienced in airway management performed endotracheal intubation in an airway manikin (SimMan, Laerdal, Kent, UK) with three different airway conditions. We evaluated the success rate using the Bonfils (Karl Storz, Tuttlingen, Germany) or the Macintosh laryngoscope, the time needed for securing the airway, and subjective rating of both techniques.
Results:
In normal airway all intubations were successful using laryngoscope (100%) vs. 82% using the Bonfils (p < 0.05). In the scenario "tongue oedema" success rate using the Macintosh laryngoscope was 67% and 83% using the Bonfils. In the scenario "decreased cervical range of motion with jaw trismus", success rate using the Macintosh laryngoscope was 84% vs. 76%. In difficult airway scenarios time until airway was secured did not differ between the two devices. Use of Bonfils was rated "easier" in both difficult airway scenarios.
Conclusion:
The Bonfils can be successfully used by physicians unfamiliar with this technique in an airway manikin. The airway could be secured with at least the same success rate as using a Macintosh laryngoscope in difficult airway scenarios. Use of the Bonfils did not delay intubation in the presence of a difficult airway. These results indicate that intensive special training is advised to use the Bonfils effectively in airway management.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...