Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography
Edward Wylegała1, Slawomir Teper, Anna K Nowińska
1Department of Ophthalmology, District Railway Hospital, Silesian Medical University, Katowice, Poland.
Purpose:
To compare anterior segment measurements and morphology of 2 optical coherence tomography (OCT) systems.
Setting:
Department of Ophthalmology, District Railway Hospital, and the Nursing Department and Social Medical Issues, Health Care Division, Silesian Medical University, Katowice, Poland.
Methods:
In normal eyes and in eyes with corneal and trabecular-iris angle disorders, the central corneal thickness (CCT), trabecular-iris angle, and angle-opening distance at the nasal and temporal angles were measured 3 times during 1 visit using the Visante time-domain OCT system and the RTVue-100 Fourier-domain corneal anterior module OCT system. Anterior segment morphology was assessed and compared.
Results:
Fifty-four eyes were evaluated. The mean values (+/-SD) by time-domain OCT and Fourier-domain OCT were, respectively, automatic CCT, 535 +/- 33.07 microm and 538 +/- 31.82 microm; manual CCT, 545 +/- 30.91 microm and 542 +/- 30.57 microm; nasal trabecular-iris angle, 34.7 +/- 9.5 degrees and 35.2 +/- 8.9 degrees; temporal trabecular-iris angle, 35.3 +/- 8.5 degrees and 35.5 +/- 9 degrees; nasal angle-opening distance, 435 +/- 95 microm and 444 +/- 98 microm; and temporal angle-opening distance, 443 +/- 103 microm and 452 +/- 99 microm. There was no significant difference between mean values, and they were highly correlated. On morphologic analysis, time-domain OCT had lower resolution; however, all anterior chamber structures were visible on 1 image. Fourier-domain OCT provided precise information about small areas of the anterior chamber.
Conclusion:
Fourier-domain OCT provided accurate anterior eye segment measurements that agreed with those obtained with time-domain OCT.
More Related Videos
06:15Anterior High-Resolution Optical Coherence Tomography in the Diagnosis and Therapeutic Monitoring of Ocular Surface Squamous Neoplasia
Published on: August 9, 2024
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Continuous -time Fourier Transform