Mass bias fractionation laws for multi-collector ICPMS: assumptions and their experimental verification
Juris Meija1, Lu Yang, Ralph Sturgeon
1Institute for National Measurement Standards, National Research Council Canada, 1200 Montreal Rd., Ottawa, Ontario K1A 0R6, Canada. juris.meija@nrc.ca
Abstract:
Although mass bias fractionation correction is among the most studied theoretical aspects of analytical mass spectrometry, several assumptions imbedded in the conventional correction models remain largely untested. Experimental evidence is given herein highlighting parts-per-thousand deviations from the conventional mass bias correction models which can occur with use of multicollector inductively coupled plasma mass spectrometry. Furthermore, current mass spectrometric approaches are not capable of elucidating the fractionation exponent.