Standardized Method for Measuring Collection Efficiency from Wipe-sampling of Trace Explosives
Published on: April 10, 2017
A NIST standard reference material (SRM) to support the detection of trace explosives
1Analytical Chemistry Division, National Institute of Standards and Technology, Gaithersburg, Maryland 20899-8392, USA.
Analytical Chemistry
|July 30, 2009
Summary
A new standard reference material, SRM 2905, was developed for simulating explosive residues from plastic and military explosives. This material, based on octadecylsilane-modified silica, aids in calibrating trace explosives detectors for enhanced security screening.
Area of Science:
- Forensic Science
- Analytical Chemistry
- Materials Science
Background:
- Handling plastic and military explosives generates trace residues.
- Accurate detection and quantification of these residues are critical for security.
- Existing methods require reliable standards for calibration.
Purpose of the Study:
- To develop a Standard Reference Material (SRM) simulating trace explosive residues.
- To ensure the material is suitable for calibrating trace explosives detectors.
- To provide certified concentrations of explosives for analytical validation.
Main Methods:
- Coating chromatographic supports with Composition C-4 (RDX/HMX) and TNT.
- Selecting a base material (C18 silica) based on coating efficiency, extractability, stability, and particle size.
- Quantifying explosive concentrations using liquid chromatography with UV and MS detection (LC/UV, LC/MS).
- Testing the SRM on an ion mobility spectrometry (IMS) detector.
Main Results:
- SRM 2905 comprises four materials with two nominal concentrations (0.01% and 0.1% mass fraction) of C-4 and TNT.
- Octadecylsilane-modified silica (C18) with 20-30 micrometer particle size was selected as the base material.
- Certified concentrations were determined using validated LC/UV and LC/MS methods.
Conclusions:
- SRM 2905 provides a reliable standard for simulating explosive residues.
- The developed SRM is suitable for calibrating trace explosives detection systems.
- This standard enhances the accuracy and reliability of explosives detection in forensic and security applications.
More Related Videos
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
In an NMR sample, precise measurement of the absolute absorption frequencies of nuclei is difficult. A standard internal reference compound is added, and the frequency difference between the reference signal and sample signals is measured.
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Atomic Absorption Spectroscopy: Lab
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...