Related Experiment Video
Updated: Jun 21, 2026

11:27
A Modified Sonographic Algorithm for Image Acquisition in Life-Threatening Emergencies in the Critically Ill Newborn
Published on: April 7, 2023
A new transesophageal probe for newborns
Thierry V Scohy1, Guillaume Matte, Paul L M J van Neer
1Department of Anesthesiology, Erasmus Medical Center, Rotterdam, The Netherlands.
Ultrasound in Medicine & Biology
|August 4, 2009
Summary
A new, smaller transesophageal echocardiography (TEE) probe was developed for newborns undergoing cardiac procedures. This innovation addresses the limitations of adult-sized probes, enabling critical cardiac imaging in neonates.
Area of Science:
- Medical Devices
- Pediatric Cardiology
- Echocardiography
Background:
- Current transesophageal echocardiography (TEE) probes are designed for adults.
- Existing probes are too large for neonatal use, limiting TEE application in newborns (<5 kg).
- There is a significant clinical need for miniaturized TEE probes for neonatal cardiac surgery and catheterization.
Purpose of the Study:
- To design and develop a small transesophageal echocardiography phased array probe suitable for neonatal patients.
- To overcome the size limitations of current TEE probes in pediatric cardiac care.
Main Methods:
- Development of a miniaturized TEE phased array probe.
- Probe specifications include a 5.2mm tube diameter and an 8.2-7 mm head size.
- The probe features 48 elements and operates at a center frequency of 7.5 MHz.
Main Results:
- Successful design and realization of a small-diameter TEE probe.
- The probe's dimensions are optimized for neonatal esophageal anatomy.
- A clinical evaluation study involving 42 patients was conducted (Scohy et al. 2007).
Conclusions:
- The developed small TEE probe is a viable solution for neonatal cardiac imaging.
- This innovation facilitates transesophageal echocardiography in newborns requiring complex cardiac interventions.