Interference with the operation of medical devices resulting from the use of radio frequency identification
Bryan Houliston1, David Parry, Craig S Webster
1AURA Laboratory, School of Computing and Mathematical Sciences, Auckland University of Technology, Auckland, New Zealand. bryan.houliston@aut.ac.nz
Aim:
To replicate electromagnetic interference (EMI) with a common drug infusion device resulting from the use of radio frequency identification (RFID) technology in a simulated operating theatre environment.
Method:
An infusion pump, of a type previously reported as having failed due to RFID EMI, was placed in radio frequency (RF) fields of various strengths, and its operation observed. Different strength RF fields were created by varying the number of RFID readers, the use of a high-gain RFID antenna, the distance between the reader(s) and the infusion pump, and the presence of an RFID tag on the infusion pump.
Results:
The infusion pump was not affected by low-power RFID readers, even when in direct contact. The pump was disrupted by a high-power reader at 10 cm distance when an RFID tag was attached, and by a combination of high-power and low-power readers at 10 cm distance.
Conclusions:
Electronic medical devices may fail in the presence of high-power RFID readers, especially if the device is tagged. However, low-power RFID readers appear to be safer.
Related Concept Videos
Bacterial Signaling
Issues And Trends In Healthcare Delivery System
Cost Containment
Payment for healthcare services has historically promoted adoption of costly and often unnecessary or inefficient...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences