Left ventricular mass from gated SPECT myocardial perfusion imaging: comparison with cardiac computed tomography

Tochi M Okwuosa1, Chetan V Hampole, Javid Ali

  • 1Department of Medicine, Section of Cardiology, University of Chicago, Chicago, IL 60637, USA.

Insights

Gated SPECT myocardial perfusion imaging (MPI) software provides reproducible left ventricular mass (LVM) measurements, but these values are not directly substitutable for CT angiography (CTA) due to systematic over/underestimations.

Area of Science:

  • Cardiology
  • Medical Imaging
  • Quantitative Analysis

Background:

  • Increased left ventricular mass (LVM) is linked to adverse cardiac events.
  • Quantitation of LVM using gated SPECT MPI software lacks extensive validation and clinical application.
  • Current gated SPECT software requires comparison against a reference standard for accurate LVM assessment.

Purpose of the Study:

  • To compare Left Ventricular Mass (LVM) measurements from two gated SPECT software packages (4D-MSPECT and QPS) against CT angiography (CTA) as a 3D reference standard.
  • To evaluate the clinical utility and accuracy of SPECT-derived LVM in comparison to CTA.
  • To determine the reproducibility and agreement between LVM quantitation methods.

Main Methods:

  • 56 patients referred for MPI and cardiac CTA underwent LVM measurement by two blinded observers.
  • Dual-isotope or low-dose/high-dose Tc-99m-tetrofosmin rest/stress MPI and cardiac CTA were utilized.
  • Interobserver reproducibility and correlation analyses, including Bland-Altman analysis, were performed.

Main Results:

  • Post-stress MPI values showed better correlation with CTA than rest values.
  • Both 4DM and QPS demonstrated high reproducibility (r = 0.99).
  • SPECT LVM values showed moderate correlation with CTA (r = 0.74 for 4DM, r = 0.72 for QPS), but Bland-Altman analysis revealed significant overestimation of lower LVM and underestimation of higher LVM values compared to CTA.

Conclusions:

  • Gated SPECT and CTA provide reproducible LVM measurements.
  • While mean SPECT LVM values are comparable to CTA, systematic biases necessitate caution.
  • SPECT LVM values require mathematical correction for accurate substitution of CTA-derived values.
Abstract